1. Main technical parameters

Model

MYLH-30E

MYLH-50E

MYLH-100E

load

30kN

50kN

100kN

Hydrogen pressure (MPa)

20, 35

20, 35, 50, 70, 100

35, 50, 70, 100, 140

Test temperature

normal temperature

(-50~100)℃

Test function

slow tensile test

Slow tensile test, crack growth rate test, fracture toughness test, low cycle fatigue test

Loading strain rate

(1/s~1x10-8)/s

(10~1x10-7)/s

Test frequency

——

(0.1~1)Hz

Test waveform

——

Sine wave, triangle wave

Test force range

1%~100%F.S

Relative error of test force indication

±0.5%

Maximum moving range of loading head

80mm

Loading head displacement measurement accuracy

±0.5%

Loading head displacement resolution

0.0125µm

Deformation measurement range

0~30mm

Deformation measurement resolution

1µm

Deformation measurement accuracy

±0.5%

Below -40℃, perform lower pressure (35MPa) test.


2. Functional features

l Direct measurement of the deformation of the gauge length section in the autoclave;

l Use an autoclave internal force sensor to directly measure the tensile load of metal materials, eliminating measurement errors caused by sealing friction resistance;

l Self-designed balancing device eliminates the impact of pressure in the kettle on the sample;

l Multiple innovative combined sealing technologies to prevent hydrogen leakage;

l Professional material selection to prevent stress corrosion cracking caused by hydrogen environment;

l Systematic design, in line with regulatory safety requirements;

l Multiple testing to ensure safety;

l Provide complete solutions for hydrogen test projects;

l Supports sample control process design to facilitate test method expansion;

l Support slow strain loading control;

l Support automatic and manual test log recording;

l Supports recovery function after test suspension;

l Support test recovery function after power failure;

l Supports retention of collected data after power failure;

l Support long-term continuous testing of more than 30 days;

l Test data can be exported to EXCEL and ACCESS formats;